NetScanner™ System (9016, 9021, & 9022) User’s Manual

ii

dd

Description

‘q’ read

index

www.PressureSystems.com

3 - 71

0B

00

01

Enable Automatic Shifting of Calibration Valve during
Calculate and Set Offsets (‘h’) command (default). (9016
only)
Disable Automatic Shifting of Calibration Valve in ‘h’.
User will manually control calibration value. (9016 only)

0C

00

01

Set Cal Valves to RUN or LEAK Position (default)
— choice made by ii=12. (9016 only)
Set Cal Valves to CAL/RE-ZERO or PURGE Position
— choice made by ii=12. (9016 only)

0B

see

chart

below

0D-0E

--

Reserved for factory use

0F

00
01

Disable periodic Thermal Coefficient Update task.
Enable periodic Thermal Coefficient Update task
(default).

10

01-20

Set Number of A/D Samples to Average. (default = 8).
Valid values are 1, 2, 4, 8,10, 20 hex: (1, 2, 4, 8,16, 32
decimal)

05

11

--

Reserved for factory use

12

00

01

Set Cal Valves to RUN or CAL/RE-ZERO Position
(default) — choice made by ii=0C. (9016 only)
Set Cal Valves to PURGE or LEAK Position — choice
made by ii=0C. (9016 only)

0B

see

chart

below

13

00
01

Use Static IP Address Resolution (default)
Use Dynamic IP Address Resolution (RARP/BOOTP)
(Results in immediately becoming the module’s new
power-on default)

06

14

00
01

02

Disable Host Response/Stream Back-Off Delay (default).
Enable Host Response/Stream Back-Off Delay as low-
order byte of Ethernet Address(converted to decimal value
* 20 µsec.).
Enable Host Response/Stream Back-Off Delay specified
per eeee as decimal value (* 20 µsec.)

07