The coefficients of internal DH200 transducers used in the Model 9016 are selected
with array indexes aa=01 through 10 (hex). Coefficients of external 9400
transducers used by Models 9021 and 9022 (digitally compensated) are selected
with array indexes aa=01 through 0C (hex). All valid coefficient indexes (for each
of these arrays) are listed in the following table:

cc

Transducer Coefficient Description

Datum Type

00

Re-zero Cal Adjustment (offset) term

(Note 1)

FLOAT

01

Span Cal Adjustment (gain) term

(Note 2)

FLOAT

02

Dynamic EU Conversion coefficient c0

(Note 3)

FLOAT

03

Dynamic EU Conversion coefficient c1

(Note 3)

FLOAT

04

Dynamic EU Conversion coefficient c2

(Note 3)

FLOAT

05

Dynamic EU Conversion coefficient c3

(Note 3)

FLOAT

07

User Defined Field

(Note 4)

INTEGER

09

Transducer Manufacturing Reference Number (Note 5)

INTEGER

0A

Transducer Full-Scale Range Code (See Appendix F) (Note
5)

INTEGER

4D

Pressure Voltage Gain Index

(Note 5)

INTEGER

4E

Temperature Voltage Gain Index

INTEGER

(Note 1)

Related command ‘w08’ can be used to download the offset term to the sensor’s
non-volatile memory (digitally or non-digitally compensated sensor).

(Note 2)

Related command ‘w09’ can be used to download the gain term to the sensor’s non-
volatile memory (digitally or non-digitally compensated sensor).

(Note 3)

Related command ‘w27’ can be used to download coefficients to the 9400 sensor’s
non-volatile memory (9021/9022 non-digitally compensated ONLY).

(Note 4)

Data is immediately stored to the sensor’s non-volatile memory.

(Note 5)

Data is immediately stored to the sensor’s non-volatile memory (9021/9022 non-
digitally compensated ONLY).