4.3.2. Span Calibration Summary

Following is a simple, step-by-step procedure for executing a “full scale” span calibration of a 9016,
9021, or 9022 Intelligent Pressure Scanner. For Models 9021 and 9022, skip the steps that
manipulate valves
, as denoted with a double asterisk (**). It is assumed that all channels in the unit
are of the same full scale pressure range. Optional commands are shown within brackets [ ].

Description

9016 Command

9021/9022

Command

Ensure that valves are in RUN/CAL mode (default).

[w1200]

**

… normal data acquisition

Perform Re-zero calibration

See Section 4.2.2

Place the module calibration manifolds into the CAL
position if this is the desired pressure application input.
The RUN position may be a better choice for modules
with transducers having different ranges.

[w0C01] for CAL

[w0C00] for RUN

**

Apply exact full scale pressure to appropriate module
CAL and CAL REF inputs [or optionally to RUN
inputs].

Delay for settling of pneumatic inputs

Verify that measured data reads near expected full scale

[rFFFF0]

[r0FFF0]

Instruct module to calculate new gain coefficients for all
16 channels

ZFFFF

Z0FFF

Place calibration manifold back into the RUN position

w0C00

**

Store new gain coefficients to transducer non-volatile
memory

w09

… continue normal data acquisition