A detailed step-by-step procedure follows that includes the specific 9021/9022 commands to be sent
to the module under test.

DESCRIPTION

9021/9022 COMMAND

Set AD averaging to 32 to minimize reading jitter

w1020

Turn off front-end calibration function (9022 ONLY)

w2200 00

Set all channels to gain of 1 (or 20, 45, 90)

1

w1500
(or w 150 1, w 150 2, or w1503 for gain of 20,45, or 90)

Read current G=1 reference coefficient from
9021/9022 in ASCII floating point format

2

u01104
(u01 18 fo r 90 22 te mpe rature r eferen ce co efficient)

Ap ply 4.5 00 V DC to 90 21/9 022 chann el 1

**

Read mea sured voltage from channel 1 in ASCII
floating point (data format 0)

r00010
(t00010 for temperature voltage)

Calculate difference between actual applied voltage
and 9021 /9022 reported vo ltage

**

Download new reference coefficient for G=1 in ASCII
floating point format

3

v0110 4 x.xxxx (x.xxxx is new coefficient)
(v01118 x.xxxx for 9022 temperature reference
coe fficient)

Store coefficient to 9021/902 2 FLAS H mem ory

w07

Reboot module to force new coefficients to take effect

**

1

The gain control commands for the 9021/9022 are summarized in the following table:

Gain

9021/9022

Command

1

w1500

20

w1501

45

w1502

90

w1503

2

Commands to read the respective gain coefficient commands are summarized in the following

table.