When performing any type of maintenance of NetScanner

System components, the

following guidelines and precautions should always be followed:

!

Verify that the work area and technicians are properly grounded to prevent damage to
electronic components due to electrostatic discharge.

!

Ensure that all electrical and pneumatic connections have been removed from the
module.

!

Ensure that the work area is free of dust and other possible contaminants that may
affect the high tolerance machined parts (and pneumatic seals, if model has an integral
manifold).

!

Care must be taken to prevent contaminants from reaching O-ring surfaces. If O-ring
surfaces require cleaning, use a lint-free applicator with acetone to remove dirt and
lightly lubricate the O-ring surface with Krytox® provided in the maintenance kit.

! Never use sharp objects to cut tubing from the bulged tubes. The tiny scratches left on

the tubes could cause leaks.

NOTE: PSI offers an optional
maintenance kit (PSI P/N
S901-0200000000
) with the
lubricants included.

In the process of performing general maintenance on a module and in printed-circuit board
replacement, the following tools may be required:

!

3/32" and 5/64" Allen-head screwdrivers,

!

a 3/16" hex wrench,

!

a needle nose tweezers,

!

a Phillips-head screwdriver,

!

a small adjustable wrench,

!

a tube of Krytox

®

lubricant (PSI P/N 42-06-KRYX),

!

a spray bottle of silicone liquid lubricant (PSI P/N 41-06-Silicone).