3 options – Pressure Systems 9022 User Manual

Page 15

Advertising
background image

Pressure Systems, Inc.

NetScanner™ System (9016, 9021, & 9022) User’s Manual

www.PressureSystems.com

1 - 4

!

Measurement flexibility — each 9021/9022 module is capable of measuring
general purpose voltage signals on any channel not populated with a 9400-type
transducer. Full-scale ranges of ±50, ±100, ±250 and ±4500 mV are supported
through programmable gain amplifier circuitry.

!

Ease of use — modules have simple command sets and provide engineering units
output. They may interface directly to a desktop or laptop computer or they may
be interconnected into a large network controlled by many types of host
computers.

!

Connectivity — use of industry-standard communications network protocols to
control and read data from NetScanner

System modules allows distribution to

the point of measurement and ensures compatibility with third party hardware and
software.

1.3

Options

1.3.1

Pressure Ranges

Model 9016 contains sixteen (16) DH200 transducers. These transducers are available with full
scale pressure ranges from 10" H

2

O (inches of water column) to 750 psid (2.5 kPa to 5200 kPa).

Transducers with different pressure ranges may be combined in a single module.

Models 9021 and 9022 can attach up to twelve (12) Series 9400 or third party external all-media
transducers. The 9400 gauge-type transducers are available with full-scale pressure ranges from
5 psi to 10,000 psi (35 kPa to 69000 kPa). The 9401 absolute-type transducers are available
with full-scale pressure ranges from 15 psia to 10,000 psia (105 kPa to 69000 kPa). The 9402
wet-wet differential type transducers are available with full scale ranges from 5 psi to 250 psi (35
kPa to 1725 kPa). Transducers with different pressure ranges may be attached to a single
module.

Please consult the Sales Department at Pressure Systems for availability of other pressure ranges
(1-800-678-SCAN (7226)).

1.3.2

Manifolds and Pressure Connections

Model 9016 sixteen-channel Intelligent Pressure Scanners are available with a true differential or
common reference pneumatic manifold, and have a standard purge and leak check manifold.
They are available with standard 1/8" or optional 1/16" and 1/4" compression fittings. All
fittings

Advertising