NetScanner™ System (9016, 9021, & 9022) User’s Manual

www.PressureSystems.com

3 - 74

The Valve Position indexes (ii=12 and ii=0C) (for Model 9016 only) each have two states (00/01)
that when combined provide four (4) possible states of the C1/C2/C3/C4 internal valves. This
“Logical Rotary Switch” with four (4) positions is summarized in the following chart:

C1 Energized

C2 Not

(0C=01)

C2 Energized

C1 Not

(0C=00)

C3 Energized

C4 Not

(12=01)

PURGE

position

LEAK/CHECK

position

C4 Energized

C3 Not

(12=00)

CAL/RE-ZERO

position

RUN

position

Example:

!

Send TCP/IP commands to 9016 module (via its connected socket) setting the
calibration valve to the CAL (or Re-Zero) position:

w1200

"

(

Set RUN/CAL valve position

)

w0C01

(Set CAL position)

Responses (both commands):

“A”
“A”

NOTE:

If the programmer “knows” that the module is already in the RUN/CAL
valve mode, the first command above is optional. For more information, see
Figures 4.1 through 4.4 in Chapter 4.