READ HIGH-SPEED DATA (Command ‘b’)

Purpose:

Returns the most recent scanned/averaged data from all channels of the module as
fast as possible. Data is returned directly in its internal (IEEE single-precision float)
binary form (as per implied format 7). It is used as a faster alternative to the Read
High-Precision Data
(‘r’) command, since ‘b’ does not have to parse the position
or format parameters, nor does it have transform or encode the internal data into any
other format when the response is generated.

Command

b

‘b’ is the command letter

Response

aaaabbbbcccc..rrrr

each 4-byte datum (e.g, ‘aaaa’) is a non-human readable 32-bit (4-byte) big-
endian value (format 7) representing an IEEE single-precision internal float
value.

Description: Returns data for all of the module’s channels, in order highest channel number to

lowest channel number. Thus for a Model 9016, channel #16 will always be the first
4-byte (32-bit binary, big-endian, IEEE floating-point) value (‘aaaa’) sent in the
response. For Models 9021 and 9022, channel #12 will be first. It is followed by
similar values for lower numbered channels.

Unless the EU conversion scalar is altered, the returned data will be in units of psi.

Example:

!

Send command to a module (via its “socket” connection) to return data as fast
possible:

b

Data from the most recent scan of all the module’s channels are returned in pure
binary form, 4-bytes per channel (big endian):

aaaabbbbcccc .. rrrr

Note that this response is not shown within quotes “ ” since it is not a valid ASCII
character string.