Pressure Systems 9022 User Manual

Page 74

Advertising
background image

Pressure Systems, Inc.

NetScanner™ System (9016, 9021, & 9022) User’s Manual

www.PressureSystems.com

3 - 44

NOTE: Selecting too many other data groups will compromise module

performance.

** NOTE: This status field (0001) cannot be specified for Models 9016, 9021, or
9022. However it is shown should the capability be added to future firmware
versions. Currently, only Models 9816 and 903x can return Valve Position status in
their streams.

bbbb (hex)

data selected for inclusion in each stream packet

0001 **

Enable Valve Position Status (reserved for future use)

0002

Enable DH Temperature Status (see bit map below)

0010

Enable Pressure EU Data (default if “05” never executed
after “00”)

0020

Enable Pressure A/D Counts

0040

Enable Pressure Voltages

0080

Enable DH Temperature EU Data (degrees C)

0100

Enable DH Temperature A/D Counts

0200

Enable DH Temperature Voltages

Any DH Temperature Status datum is delivered as a two-byte binary bit map (16-
bit, big endian) with each bit representing the status of DH #16 through DH #1
respectively. A bit value of 0 (zero) indicates the DH is operating within its
specified limits. A value of 1 (one) indicates the DH is outside its specified limits.

Bit #

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

Chan #

16

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

Binary

1

0

0

0

0

0

0

0

0

0

0

0

0

0

0

1

Hex

8

0

0

1

The above example indicates that Channels 1 and 16 are operating outside the
specified temperature limits.