Pressure Systems, Inc.

NetScanner™ System (9016, 9021, & 9022) User’s Manual

www.PressureSystems.com

3 - 15

RESET (Command ‘B’)

Purpose:

Instructs the module to reset internal operating parameters, and to set all internal
control variables to their default “reset” state (see description below). The current
TCP/IP socket connection will remain open. Execution after a power off/on cycle
is optional (unnecessary).

Command

B

‘B’ is the command letter.

Response

A

‘A’ is the acknowledge letter.

Description: The module returns to the following “reset” states if this command is executed:

!

Re-zero correction (offset) terms are set to the last values stored in
transducer memory.

!

Span correction (gain) terms are set to the last values stored in
transducer memory.

!

Calibration Valve is set to the RUN Position (9016 only).

!

Number of Samples for Data Averaging is set to last value stored in
non-volatile memory (factory default = 8).

!

Any autonomous host data delivery streams defined by ‘c’ sub-
commands are reset (undefined).

!

The Multi-Point Calibration function defined by ‘C’ sub-commands
is reset (undefined) if in progress.

Example:

!

Send TCP/IP command to a module (via open socket) to reset defaults:

B

Read following response:

A