Pressure Systems 9022 User Manual

Page 119

Advertising
background image

Pressure Systems, Inc.

NetScanner™ System (9016, 9021, & 9022) User’s Manual

www.PressureSystems.com

4 - 9

Following is a simple, step-by-step procedure for executing a specified-value span calibration of
a 9016, 9021, or 9022 Intelligent Pressure Scanner. For Models 9021/9022, skip the steps that
manipulate valves
as denoted by the double asterisk (**). For the purposes of this example, it will
be assumed that an upscale pressure of 14.9800 psi is available from a dead weight tester for the
calibration of 15 psi internal transducers. All sixteen channels are 15 psi full scale.

Description

9016 Command

9021/9022 Command

Ensure that valves are in RUN/CAL mode
(default)

[w1200]

**

… normal data acquisition

Perform Re-zero calibration first

See Section 4.2.2

Place the module calibration manifolds into
the CAL position if this is the desired
pressure application input. The RUN position
may be a better choice for modules with
transducers having different ranges.

[w0C01] for
CAL

[w0C00] for
RUN

**

With a deadweight tester, apply 14.9800psi
to the appropriate module CAL and CAL
REF
inputs [or optionally to RUN inputs]

Delay for settling of pneumatic inputs

Verify measured data reads near expected full
scale

[rFFFF0]

[r0FFF0]

Instruct module to calculate new gain
coefficients for all 16 channels

ZFFFF 14.98

Z0FFF 14.98

Place calibration manifold back into the RUN
position

w0C00

**

Store new gain coefficients to transducer non-
volatile memory

w09

… continue normal data acquisition