NetScanner™ System (9016, 9021, & 9022) User’s Manual

www.PressureSystems.com

4 - 16

The above process is repeated for each of the four (4) 9021/9022 gain settings (or selected gain
settings as needed for your application).

Calculation of the new reference coefficient is performed as follows:

CoefOld = current reference coefficient value

CoefNew = new/corrected coefficient value

Vapplied = actual applied test voltage

V9021 = reported voltage measured by the 9021/9022

CoefNew = CoefOld * [1 + (Vapplied - V9021) ÷ Vapplied]

Example:

CoefOld:

9021/9022 Gain=1 current reference coefficient is 0.9800

Vapplied:

4.004 VDC is applied to 9021/9022 channel 1 from a precision voltage
source

V9021:

9021/9022 reports 3.965 V measured at channel 1

CoefNew = 0.9800 * [1 + 4.004 - 3.965) ÷ 4.004]

CoefNew = 0.98955