DOWNLOAD INTERNAL COEFFICIENTS (Command ‘v’)

Purpose:

Downloads one or more internal coefficients to the module.

Command

vfaacc[-cc]dddd..dddd

‘v’ is the command letter.
f’ is the format field.
aa’ is the array index field.
cc[-cc]’ is coefficient index [or contiguous range].
dddd’ are the datum field(s) each with a leading
space.

Response

A

‘A’ is the acknowledge letter.

Description: The 1-character format field (f) is a single decimal digit that defines the format of

each coefficient to be downloaded in the command’s datum ( dddd) fields, with each
datum preceded by a space character. Most coefficients have a floating point datum
type (f=0-1), while others have an integer datum type (f=5). Sending a datum in the
improper format will result in an “N08" error response. Valid format types are
shown in the following table:

f

converts each datum parameter value (‘ dddd’) from..

max. char.

0

1-10 digit signed decimal “ [-xxx]x.[xxxxxx]”

to

single binary float

13

1

8-digit hex “ xxxxxxxx”

to

single binary float

9

5

8-digit hex “ xxxxxxxx”

to

long binary integer

9

The 2-character array index field (aa) is a hexadecimal value selecting a particular
internal coefficient array to receive the downloaded data. The first array index
(aa=01) refers to channel 1's transducer, the 16th (aa=10) refers to channel 16's
transducer. Finally, the last array (aa=11) refers to a special global array.

A single 1- or 2- character coefficient index field (c or cc) is a hexadecimal value
that selects a particular coefficient within the specified array. Multiple contiguous
coefficients may be specified by using a coefficient index “range” specified by
adding a hyphen (negative sign) between two such indexes.