Level, Level -4 – RIGOL DSG3000 Series User Manual

Page 132

Advertising
background image

RIGOL

Chapter 6 Specifications

6-4

DSG3000 User’s Guide

Level

Setting Range

Specification level range Setting range

Maximum output

level

9kHz ≤ f <100kHz +7dBm

+10dBm

100kHz ≤ f <1MHz +13dBm

+15dBm

1MHz ≤f ≤3GHz

+13dBm

+25dBm

3GHz <f ≤6GHz

+13dBm

+20dBm

Minimum output

level

9kHz ≤ f <100kHz -110dBm

-120dBm

100kHz ≤ f ≤ 6GHz -130dBm

-140dBm

Setting resolution

0.01dB

Absolute Level Uncertainty

[1]

Level

uncertainty

+13 to

-60dBm

-60 to

-110dBm

-110 to -130dBm

9kHz ≤f<100kHz ≤ 0.5dB (typ.) ≤ 0.7dB (typ.)

100kHz ≤f≤3GHz ≤ 0.7dB,

≤ 0.5 (typ.)

≤ 0.9dB,

≤ 0.5 (typ.)

≤ 0.7dB (typ.)

3GHz < f ≤ 6GHz ≤ 0.9dB,

≤ 0.5 (typ.)

≤ 1.1dB,

≤ 0.5 (typ.)

≤ 0.9dB (typ.)

VSWR

[2]

1MHz ≤ f ≤ 6GHz < 1.8 (typ.)

Level Setting

Setting time

ALC state on, frequency fixed,

temperature range: 20℃ to
30℃

≤ 5ms (typ.)

Uninterrupted level

setting range

ATT fixed mode, ALC state on,

level range -110dBm to

+13dBm

>20dB (typ.)

Max. Reverse Power
Max. reverse power Max. DC voltage

50V

1MHz < f ≤ 6GHz

10W

NOTE: [1] ALC state: on or auto mode, 20℃ to 30℃

[2] In 50Ω system, typical, level ≤-10dBm, ATT auto mode

Advertising