Zero type, Zero, Frequency mode – RIGOL DSG3000 Series User Manual

Page 85: Frequency, Zero type -45, Zero -45, Frequency mode -45, Frequency -45

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DSG3000 User’s Guide

2-45

Zero Type

You can execute the power meter zeroing to reduce the zero measurement deviation

and the effect of noise to improve the RF power measurement accuracy.

Press Zero Type to select “Int” or “Ext” zero mode.


Internal zero

When selecting this mode, the power sensor can still be connected with the RF

signal during the zeroing process. The instrument uses the internal zero by

default after powered-on.

External zero

When selecting this mode, you must remove the RF signal to be measured

before zeroing the power sensor.

Zero

Press Zero and the instrument will perform the power meter zeroing operation

according to the “zero type” selected.

Frequency Mode

Select the configuration mode of the measurement frequency.

Press Freq Mod to select “Manul” or “Auto” frequency mode.
 Manul: in this mode, you can manually set the measurement frequency using

Frequency softkey.

 Auto: the instrument configures this frequency according to the current output

frequency of RF signal generator automatically.

Frequency

Press Frequency to set the measurement frequency manually.

Tips
When the “Auto” frequency mode is selected, the frequency mode will switch

to the “Manul” mode automatically after you set the frequency value via

Frequency.