Step sweep – RIGOL DSG3000 Series User Manual

Page 53

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DSG3000 User’s Guide

2-13

Step Sweep

1. Select the step sweep mode

Press SWEEPType to select “Step”. At this point, the RF signal generator
performs a step sweep according to the current settings.

2. Set the sweep parameters

Press SWEEP and use the page up/down key

to open the 2/3 menu page;

then, press Step Swp to set the start frequency, stop frequency, start level,

stop level, sweep points and so on.

 Start Frequency

Press Start Freq, use the numeric keyboard to input the start frequency

value and select the desired unit from the pop-up unit menu or unit keys.

You can also press Enter to select the default unit (Hz).

 Stop Frequency

Press Stop Freq, use the numeric keyboard to input the stop frequency

value and select the desired unit from the pop-up unit menu or unit keys.
You can also press Enter to select the default unit (Hz).

Explanation
The start frequency and stop frequency are the frequency upper and

lower limits of the frequency sweep.
 When the sweep direction is forward, the RF signal generator

sweeps from the start frequency to the stop frequency.

 When the sweep direction is down, the RF signal generator sweeps

from the stop frequency to the start frequency.

When the “Start Freq” or “Stop Freq” is modified, the RF signal

generator will restart the sweep and output from the specified “Start

Freq” or “Stop Freq”.

 Start Level

Press Start Lev, use the numeric keyboard to input the start level value

and select the desired unit from the pop-up unit menu or unit keys. You can
also press Enter to select the default unit (dBm).

 Stop Level

Press Stop Lev, use the numeric keyboard to input the stop level value and

select the desired unit from the pop-up unit menu or unit keys. You can also
press Enter to select the default unit (dBm).