RIGOL DSG3000 Series User Manual

Page 58

Advertising
background image

RIGOL

Chapter 2 Front Panel Operations

2-18

DSG3000 User’s Guide

 Bus Trigger

After selecting “Bus” trigger, if the sweep mode is set to “Cont”, the

instrument starts to sweep a point each time the “*TRG” command is sent.

If the sweep mode is “Single”, press Single to meet the single sweep

condition. At this point, the instrument starts to sweep a point and stops

after the sweep period is finished each time the “*TRG” command is sent.

 External Trigger

In external trigger mode, the RF signal generator accepts the trigger signal

input from the [TRRIGER IN] connector (as shown in the figure below) at

the rear panel. If the sweep mode is set to “Cont”, the instrument starts to

sweep a point each time a TTL pulse signal with specified polarity is

received. If the sweep mode is “Single”, press Single to meet the single

sweep condition. At this point, the instrument starts to sweep a point and

stops after the sweep period is finished each time a TTL pulse signal with

specified polarity is received.

To specify the polarity of the TTL pulse signal, press Trig Slope to select

“Pos” or “Neg” and the default is “Pos”.

Explanation

When executing the sweep operation, the priority order of the conditions is

Single Sweep  Trigger Mode  Point Trigger Mode. For example, when

both the trigger mode and point trigger mode are set to “Key” trigger:
 In “Cont” sweep mode, press Trigger to meet the trigger mode of the

whole period and then press Trigger again to meet the trigger mode
of points within the sweep period; then, the instrument starts to

sweep.

 In “single” sweep mode, press Single to meet the single sweep

condition and then press Trigger twice to meet the sweep period
trigger mode and point trigger mode respectively. Then, the

instrument starts to sweep.

Advertising