Chapter 2 Front Panel Operations

2-8

DSG3000 User’s Guide

1. Set the flatness calibration state

Press LEVEL and use the page up/down key

to open the 2/2 menu page;

then, press FlatnessSwitch to set the flatness calibration to “On” or “Off”.

When the flatness calibration is enabled, the UF label will be displayed in the

amplitude area in the user interface.

2. Create the flatness calibration list

Press LEVELFlatnessCal List to enter the editing interface of flatness
calibration list.

SN

Row number. You can use Insert and Delete to add and delete

rows.

Freq Frequency. The flatness calibration list can contain the amplitude

calibration values of up to 6000 frequency points. The 6000

frequency points can locate in any frequency segment within the

maximum frequency range available.

You can save calibration lists applicable to different test settings or

frequency ranges and recall different calibration list when required.

Level Amplitude calibration value. When calibrating the amplitude of the

current frequency point, only the actual output amplitude will be

affected and the amplitude display value will not change.

 Insert

Press Insert to insert a row with the same calibration value below the

current row. At this point, the number of rows of the flatness calibration list

increases by one.

 Delete

Press Delete to delete the calibration value of the row currently selected.

At this point, the number of rows of the flatness calibration list reduces by

one.

 Goto Row

Press Goto Row to select the desired row for editing in the pop-up menu.

It is usually used to modify the calibration value of the specified row.

N: press n and use the numeric keyboard to input the desired row number

to select that row. The row number input should not be greater than the

total number of rows of the current list.

Top: press Top to select the first row of the current list.

Mid: press Mid to select the row at the middle of the current list.

Bottom: press Bottom to select the last row of the current list.


When the desired row is selected, press

to return to the previous

menu to modify the calibration value of the row currently selected.