Baseband signal – RIGOL DSG3000 Series User Manual

Page 75

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DSG3000 User’s Guide

2-35

Baseband Signal

1. Baseband Output

Press I/QBaseBandSwitch to select “On” or “Off”.

 On: enable the baseband output function. The RF signal generator outputs

the I (In-Phase) and Q (Quadrature Phase) components of the I/Q

modulation from the [I OUT] and [Q OUT] connectors on the rear panel.

 Off: disable the baseband output function.

2. Baseband Level

Press I/QBaseBandLevel to set the amplitude of the baseband output
signal.

 Use the numeric keyboard to input the amplitude value and select the

desired unit from the pop-up unit menu or unit keys. You can also press

Enter to select the default unit (dBm).

 The amplitude units available are V, mV, μV, nV and dBm.
 You can press the left/right direction keys to enter the parameter editing

state. Then, move the cursor to the specified digit and use the up/down

direction keys or knob to modify the parameter value.

 You can also use the up/down direction keys or knob to modify the

amplitude at the current step.

 Press Step to set the step.

3. ARB

Press I/QBaseBandARB to enter the store and recall interface. At this
time, the message “You can load the ARB files” is displayed in the user interface.
Press Enter or the right direction key to unfold the current directory and select
the ARB file (namely wavetable file; the format is *.arb) to be loaded. Then

press Recall to load this file. The message “File loading succeeded” is displayed

after the file is loaded. For more information about how to load a file, please

refer to Store and Recall.

Tips
The loadable ARB files include the default wavetables of the instrument and

the wavetables which are downloaded to the DSG3000 using the Ultra IQ

Station PC software. For more information about how to download the ARB

file, please refer to

Ultra IQ Station_Help Document

.