7. External Trigger Input Connector

When the trigger type of SWEEP or “Swp-Sine”
of LF output is “Ext”, this connector is used to

input the external trigger signal. You can set the

polarity of trigger signal by pressing Trig Slope

to select “Pos” or “Neg”

8. Signal Valid Output Connector

When the RF output frequency or amplitude is

modified, after a certain response and processing

time, the internal circuit of the instrument

outputs RF signal with specified frequency and

amplitude via the RF output connector at the

front panel. During this process, the [SIGNAL

VALID] connector outputs a pulse sync signal,

indicating that the RF output signal is valid.

High Level (+3.3 V): indicates that the RF

signal is in configuration;

Low Level (0 V): indicates that the RF signal

is stable (namely, the signal is valid).

3.3V

0V

Power-on Time

RF Valid

Configuration

Time

Configuration

Time

RF Valid

… …

V

T

9. Sweep Output Connector

When the RF sweep function is enabled, the

[SWEEP OUT] connector outputs a signal (0 V

to +10 V) while the RF output connector at the

front panel outputs the sweep signal according to

the current setting. Each sweep corresponds to

an output cycle.

When the sweep type is step, this output

signal is related to the choice of sweep shape

which can be set to “Triangle” or “Ramp”.

When the sweep type is list, the default of

this output signal is ramp.