Amprobe, Warning – Amprobe DMIII-Multitest Power-Quality-Recorder User Manual

Page 16

Advertising
background image

AMPROBE

DMIII MULTITEST

EN-14

6.1.1. Calibrating the test leads ("CAL" Mode)

1. Connect the black and yellow test leads to T1 and T4 input terminals

respectively.

T1 T2 T3 T4

I1 I2 I3

(V1) (V2) (V3)

(COM)

Connection of instrument terminals during calibration procedure.

2. If the test leads supplied with the instrument are not long enough for the

measurement you can extend the black cable.

3. Short-circuit the measuring cable ends making sure that the conductive parts of the

alligator clips make good contact with each other (see previous picture).

4. Press the F2 key. The instrument performs the calibration.


WARNING

Never disconnect the test leads when the message "MEASURING" is
displayed.


LOW

05.06.01

A numerical value

----

R+ R-

----

----

---mA ---mA

AUTO 0.11

in this field means
that the instrument
has

been

calibrated;

this

value remains on
the display for
any

further

measurement
even though the
unit is switched off
and on again.

FUNC CAL

5. At the end of the test the result is stored and used as OFFSET (that is to say

that it is subtracted from any continuity test carried out) for all the
subsequent measurements.



Note:

The instrument performs the calibration only if the resistance of the test leads is
lower than 5

.