Amprobe – Amprobe DMIII-Multitest Power-Quality-Recorder User Manual

Page 73

Advertising
background image

AMPROBE

DMIII MULTITEST

EN-71

10.2.2. Rotary Switch during a recording
If You move the rotary switch during a recording the following screen will appear:






Recording

Recording


This page means that a recording is running but the actual rotary switch position doesn't
correspond to this.

The instrument will continue to record.



10.3. STOPPING A RECORDING / ENERGY MEASUREMENT

The instrument uses a protective routine to avoid the risk of being disturbed or interrupted
during a recording or an energy measurement. Once a recording or a direct energy
measurement (see paragraph 7.7.2) has been started (with the option PASSWORD
enabled), after about 3 minutes from the last key pressure or switch rotation it won’t be
sufficient to press START/STOP (if a recording is running) or F2 (if an energy measuring is
running) to stop the recording, it will be necessary to insert the password.
In order to insert the password (which is not changeable), press the multifunction keys in
the following sequence (within 10 seconds):

F1, F4, F3, F2


In order to enable/disable this option see paragraph 7.1.

If a wrong password is inserted, the instrument will display an error message and will
repeat the request.

If no key is pressed after about 10 seconds the instrument returns back to the original
screen.


Advertising