6.2.4. "MAN" and "TIMER" mode faulty cases

 If

the

instrument

detects the External
Power supply adapter
the instrument will
show the message
displayed to the right.

M

05.06.01

-.- -

M

---V 15s

REMOVE POWER


MAN 500V

Disconnect

the

External

Power Supply Adapter

FUNC VNOM

 If

the

instrument

detects

a

Voltage

between

the

input

terminals higher than
15V, the instrument
does not perform the
test and displays the
screen to the right for
5 seconds.

M

05.06.01

-.-

-

M

---V 15s

VOLT IN INPUT

MAN 500V

ATTENTION: the test can't be
executed. Check that the circuit is
not energized.

FUNC VNOM

This result can't be saved


 If the instrument can't

generate the Nominal
Test Voltage it will
emit a long beep and
displays

a

screen

similar to the screen
to the right.

M

05.06.01

Insulation Resistance

1.17

M

107V 15s

MAN 500V

ATTENTION:

the

test

of

resistance R

ISO

was taken at a

voltage value lower than the
set

rated

voltage.

Low

insulation

case.

This

case

occurs under low insulation
conditions or in the presence of
capacitance on the installation.

FUNC VNOM

Test Duration