Amprobe – Amprobe DMIII-Multitest Power-Quality-Recorder User Manual

Page 17

Advertising
background image

AMPROBE

DMIII MULTITEST

EN-15

TEST LEADS

Before each measurement always assure that the
calibration is for the cables in use. During a
continuity test, if the resistance value free of
calibration (that is the resistance value less the
calibration offset value) is negative, the symbol

is displayed. Probably the calibration resistance
value stored in the instrument memory is not for
the cables in use, therefore a new calibration must
be performed.


6.1.1.1. Procedure to reset test leads calibration parameters

 To cancel calibration

parameters

it

is

necessary to perform a
calibration procedure
with a resistance of
test leads higher than
5

(for example with

open test leads). When
a cancellation is effected
the screen to the right
is displayed temporarily.

LOW

05.06.01

Message

>99.9

:

means

that

the

instrument detected a
resistance

higher

than 5

therefore it

will

proceed

with

Reset procedure.

>

99.9

R+ R-

----

----

---mA ---mA

AUTO 0.11

FUNC CAL