Mtr meas menu, Mtr menu – HP 8360 User Manual

Page 274

Advertising
background image

Mtr

Menu

Function Group

POWER

Menu Map

Description

This

accesses the meter measure softkeys.

C o r r A l l

Measures flatness correction values for all

frequency points in the flatness correction
array.

Measures a flatness correction value for the
frequency point currently in the active line of
the flatness correction array.

Corr Undef

Measures flatness correction values for all

frequency points in the flatness correction
array that have no correction values assigned.

The meter measure function uses an external HP

power meter

to automatically measure and store power correction values for the

frequency points requested.

Programming Codes

SCPI:

NONE, see Fltness Menu

Analyzer:

NONE

See Also

Flatness

Menu

“Creating and Applying the User Flatness Array,” in Chapter 1.