HP 8360 User Manual

Page 331

Advertising
background image

HP 8360

User’s Handbook

<manual

entry point> a string response that identifies the

paragraph number in the

HP 8360 Assembly-Level Repair Manual to

begin the troubleshooting procedure.

l

DIAGnostics:TEST:CONTinue

Causes the

execution to continue when paused for raw data

examination. Does nothing in other conditions.

l

DIAGnostics:TEST:DATA:DESC?

Returns the description string of the raw data examined during a

selftest. It returns

in other conditions.

l

DIAGnostics:TEST:DATA:MAXimum?

Returns the upper test limit for the raw data point examined.

Returns “0” in other conditions.

l

DIAGnostics:TEST:DATA:MINimum?

Returns the lower test limit for the raw data point examined.

Returns “0” in other conditions.

l

DIAGnostics:TEST:DATA:VALue?

Returns the raw data value for the raw data point examined.

Returns a “0” in other conditions.

l

DIAGnostics:TEST:DISable

Prevents the listed selftests from being selected. If ALL is sent then

all of the selftests are disabled. *RST causes

ALL to

execute.

l

DIAGnostics:TEST:ENABle

Enables the listed selftests to execute. If ALL is sent then all of the

selftests are enabled.

causes DIAG:TEST:ENAB ALL to execute.

l

DIAGnostics:TEST[:EXECute]

The specified

is executed. Normal instrument operation is

suspended and the instrument state is restored upon exiting the

mode.

l

DIAGnostics:TEST:LOG:SOURce

l

DIAGnostics:TEST:LOG:SOURce?

Sets and queries the source for the raw data logging. ALL specifies

that all raw data points are displayed. FAIL selects only those data

points out of the test limits. Both commands are executable in

mode. After

the setting is FAIL.

Operating and Programming Reference

Advertising