Hint – HP 8360 User Manual

Page 56

Advertising
background image

External Leveling Used With the Optional Step Attenuator

Some external leveling applications require low output power

from the synthesizer. The synthesizer automatically uncouples the

attenuator from the ALC system for

external leveling points.

Press

(

P O W E R L E V E L

).

Note the display. It shows:

0

POWER LEVEL: 0.00

For example, leveling the output of a 30

gain amplifier to a level

of -10

requires the output of the synthesizer to be around

-40

when leveled. At some frequencies this level is beyond the

range of the ALC modulator alone. If so, the LOW UNLVLED warning
message is displayed. Inserting 40

of attenuation results in an

ALC level of 0

which is well within the range of the ALC. At

20

attenuation is a better choice as it results in an ALC

level of -10

This gives a margin for AM or other functions

that vary the power level.

For optimum display accuracy and minimum noise, the ALC
level should be greater than -10

This is achieved by using

attenuation equal to the tens digit of output power. Example:
desired output power = -43

use:

40

, ALC -3

1. Press POWER

Select Set

Hint

To obtain flatness corrected power refer to “Optimizing Synthesizer
Performance, Creating and Applying the User Flatness Correction

Array,” later in this section.

Getting Started Advanced

Advertising