Sweep time, Swptime auto, Auto – HP 8360 User Manual

Page 374

Advertising
background image

Function Group

SWEEP

Menu Map

Description

This

lets you set a sweep time for frequency sweeps or power

sweeps. The sweep time range is 10 ms to

but the fastest sweep

time is constrained by the frequency span. The fastest possible sweep
can be determined automatically:

1. Press SWEEP (MENU), this reveals the sweep menu keys.

Select more

, to scroll to the next page of the sweep menu.

Select

Auto,

to set the sweep time to automatic. The

synthesizer calculates the fastest possible calibrated sweep time for

any

sweep span.

Whenever you press

the active entry area displays

the current sweep time and whether the sweep time is coupled to
the frequency span (far right hand side displays:

AUTO).

If the word

AUTO is not displayed then the sweep time auto function is off.

Programming Codes

SCPI: SWEep[:FREQ

<num>[time suffix] or

Analyzer: ST <num>

See Also

Power Sweep

“Power Level and Sweep Time Operation,” in Chapter 1.
“Programming Typical Measurements,” in Chapter 1.

Auto

Function Group

SWEEP

Menu Map

Operating and Programming Reference

HP 8360

User’s Handbook

Advertising