HP 8360 User Manual

Page 75

Advertising
background image

9. Press (PRIOR). Leave the delete menu and return to the previous

soft key menu.

10. Select Auto Fill Start

Set the first frequency in

correction table to 2

11. Auto Fill Stop

Set the last frequency in

correction table to 20

12. Auto Fill

Set the frequency increment to

every 100 MHz from 2 to 20

Setup Power Meter

13. Zero and calibrate the power meter/sensor.

14. Connect the power sensor to test port.

15. Enter and store in the power meter, the power sensor’s cal factors

for correction frequencies to be used.

Enter Correction Data into Array

16. Select

Menu

Measure

All . The power meter

is now under synthesizer control and is performing the sequence
of steps necessary to generate the correction information at each
frequency point.

If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the synthesizer is
using (address 13 is assumed). Refer to the menu map 8, System,

for the key sequence necessary to reach

Meter Adrs .

Enable User Flatness Correction

When the operation is complete, (a message is displayed) the

flatness correction array is ready to be applied to your setup.

18. Disconnect the power meter/sensor.

19. On the synthesizer, press

Where

=

l o s s

for maximum leveled power at the test

port.

20. To save the synthesizer parameters including the correction table

in an internal register, press

(n = number 1 through 8).

Reactivate the HP

8757

System Interface

21. Set the analyzer to SYSINTF ON, the analyzer and synthesizer

preset.

22. Press (RECALL)

Recall the synthesizer parameters from storage

register 1

Getting Started Advanced

Advertising