HP 8360 User Manual

Page 392

Advertising
background image

INTERRUPT: This can only be caused by an internal

processor error. Refer to the “OPERATOR’S CHECK” chapter for
instructions on contacting a qualified service technician.

INTERRUPT: This can only be caused by an internal

processor error. Refer to the “OPERATOR’S CHECK” chapter for
instructions on contacting a qualified service technician.

Number of pts must be

2 !!: This error occurs in association

with the user power flatness, auto fill number of points, feature. If
the number of points requested is less than two, you will see this
error message. Correct by entering number of points greater than or
equal to two.

OPTION NOT INSTALLED: This error occurs when the HP-IB

language switch is set to a configuration requiring a certain
firmware/hardware combination to be present in the synthesizer. See

“INSTALLATION” for information on language selection and see
“Specifications” for information on option available.

PRIV VIOLATION EXCEPTION: This can only be caused by an

internal processor error. Refer to the “OPERATOR’S CHECK”
chapter for instructions on contacting a qualified service technician.

RECALL REGISTERS LOST: This message can appear in
association with the security menu feature, memory clear. Also, a
weak, dead, or disconnected internal battery can cause this message.
Refer to the “OPERATOR’S CHECK” chapter for instructions on
contacting a qualified service technician.

REQUIRES system interface OFF: This error message

indicates that the synthesizer is connected to a network analyzer and

can not run selftest. Correct by disconnecting the system interface
cable from the synthesizer.

SPURIOUS INTERRUPT: This can only be caused by an internal

processor error. Refer to the “OPERATOR’S CHECK” chapter for
instructions on contacting a qualified service technician.

SYSTEM CONTROLLER ON BUS: This error message is

generated when an external controller is active on the HP-IB and the
synthesizer has attempted to act as the controller. Disconnect the
HP-IB interface or return the synthesizer to LOCAL operation and
repeat the request.

TOO MANY CORRECTION PTS REQUESTED: This error occurs

in association with the user power flatness menu. The maximum
number of correction points has been reached or the addition of the
points requested will exceed the maximum. The maximum number of
points available is 801.

TOO MANY LIST POINTS REQUESTED: This error occurs in
association with the frequency list menu. The maximum number of

list points has been reached or the addition of the points requested
will exceed the maximum. The maximum number of points available
is 801.

Error Messages