using (address 13 is assumed). Refer to the menu map 8, System,
for the key sequence necessary to reach

Meter Adrs .

Enable User Flatness Correction

13. When the operation is complete, (a message is displayed) the

flatness correction array is ready to be applied to your setup.

14. To save the synthesizer parameters including the correction table

in an internal register, press

(n = number 1 through 8).

15. Disconnect the power meter/sensor and press

[amber LED on). The power produced at the point where the
power meter/sensor was disconnected is now calibrated at the

frequencies and power level specified above.