1 static test (estimated battery life) – Frye FONIX 7000 User Manual

Page 109

Advertising
background image

Automated Test Sequences

101

Figure 4.2—Battery Current Test

1. mA/Freq graph
2. mA/Ampl graph
3. Static test (estimated battery life)
4. Selected test settings.

4.2.1 Static Test (Estimated Battery Life)

The Static test estimates the battery life of the hearing aid. To turn on the test:

1. Enter the Battery screen (as described above).
2. Use [F7] to set the battery size: Press [F7]. Use [∨, ∧] to select the battery

size of the hearing aid. Use [>] to complete the selection and close the pop-
up menu.

3. Use [F4] to toggle the Static Tone to On. The estimated battery life should

now be displayed below the ma/Freq graph. See Figure 4.2.

By default, this test uses a pure-tone stimulus of 1000 Hz at 60 dB SPL. The
current source selections are shown in a box below the mA/Ampl graph on the
right side of the screen. You can:

• Change the source Type by using [F3]. Toggle between Puretone and

Composite,

• Change the source Amplitude by using [∨, ∧], and
• Change the source Frequency by using [<, >]. This setting is only avail-

able when the soucre type is set to Puretone with [F3].

Advertising