Frye FONIX 7000 User Manual

Page 124

Advertising
background image

116

FONIX 7000 Hearing Aid Analyzer

• OSPL90 Sweep Meas Delay—The delay between frequencies in the

OSPL90 sweep.

• Res Crv Sweep Start Delay—The delay after the first tone is presented in

the response curve (at 50 or 60 dB SPL) sweep and before the first mea-
surement is taken.

• Res Crv Sweep Meas Delay –The delay between frequencies in the

response curve (at 50 or 60 dB SPL) sweep.

• Misc Start Delay—The delay after the first tone is presented and before

the first measurement is taken in the HFA/SPA, reference test gain, equiv-
alent input noise, and harmonic distortion measurements.

• Misc Meas Delay—The delay between frequencies in the measurements

mentioned above.

Note: ANSI 87 combines the “OSPL90” and “Res Crv” delays into the “Sweep
Start” delay settings.

AGC Settings
The purpose of the following settings is identical for each I/O frequency.

• AGC frequency—Status of the I/O test at the selected frequency. Turn the

test ON or OFF.

• I/O Start—The delay after the first tone is presented in the I/O sweep and

before the first measurement is taken.

• I/O Meas—The delay between frequencies in the I/O sweep.
• Attack Window—The amount of time the analyzer measures the attack

time of the hearing aid. Select a time that is at least twice as long as the
attack time specified by the manufacturer.

• Release Window—The amount of time the analyzer measures the release

time of the hearing aid. Select a time that is at least twice as long as the
release time specified by the manufacturer.

Note: ANSI 87 only measured I/O at 2000 Hz, and the Attack and Release win-
dow settings are not available.

Misc Settings

• Print Label—Status of the printing label. Choose ON or OFF.
• Printer—Printer used for printouts. Choose INTERNAL or EXTERNAL.

See Section 2.4.