Frye FONIX 7000 User Manual

Page 81

Advertising
background image

Basic Sound Chamber Tests

73

3.4.5 Running a Single Frequency Measurement

At times, you may want to find out what the frequency response of the hearing
aid is to a single pure-tone signal. See Figure 3.4.5.

1. Press [MENU] from the Coupler screen.
2. Use [∨, ∧] to select Static Tone under Source Settings.
3. Use [<, >] to choose SINGLE.
4. Press [EXIT] to return to the Coupler screen. The single tone will automati-

cally be running.

5. Adjust the frequency using the [<, >] buttons. Adjust the amplitude using

the [∧,∨] buttons.

Figure 3.4.5—Measuring a single frequency

3.4.6 Running a Three-Frequency Average

The ANSI standard includes some three frequency average tests. You can run
a three frequency average test (apart from the test sequence) in the Coupler
screen. See Figure 3.4.6. To do this:

1. Press [MENU] from the Coupler screen.
2. Use [∨, ∧] to select Static Tone under Source Settings.