Frye FONIX 7000 User Manual

Page 182

Advertising
background image

174

FONIX 7000 Hearing Aid Analyzer

SPECTRUM MODE AVERAGING

Spectral : off, 2,4,8 and 16.

HARMONIC DISTORTION ANALYZER

Type:

Selectable for 2nd, 3rd, Total (2nd plus 3rd), or none.

Resolution:

0.1 percent.

Reading:

Percent (%) with respect to fundamental. Single frequency read-
ings made at 100-Hz intervals from 200 through 2600 Hz. Sweep
readings made at 100 Hz intervals from 400 through 1900 Hz.

ATTACK/RELEASE TIME

Range:

1.25 to 5000 mSec.

Accuracy:

± 10% or 2 mSec + Resolution, whichever is larger.

Resolution:

2000-8000 Hz: 1.25 mSec;
800-1600 Hz: 2.5 mSec;
400, 500 Hz: 5 mSec.
100 to 300 Hz and Composite: 10 mSec.

Available Test Modes:

ANSI 96, ANSI 87

Signal Durations:

ANSI 87: 2000 mSec

ANSI 96: variable (500 to 5000 mSec)

TEST SEQUENCES

Choose ANSI or IEC as part of standard package. Additional charge for the second test sequence.
ANSI 3.22 (87/92/96)
IEC EN60118-7 1993

FEATURES

Group Delay and Phase Measurement
Battery Current Measurement
Coupler Multi-Curve
RS232 Function

AVAILABLE SIGNALS

Composite, Sine, Digital Speech (ICRA & ANSI)

OPTIONS

Real Ear Measurement (7000 Quik-Probe), 2nd Test Sequence, VA CORFIG, MZ coupler.

PRIMARY POWER

100-240 VAC 50/60 Hz. Power requirement is 50 VA.