Fitting formula tables, Appendix e, Nal-rp fitting formula calculation – Frye FONIX 7000 User Manual

Page 197

Advertising
background image

Appendix E

189

Fitting Formula Tables

The conversions used by the FONIX 7000 Test System to convert
audiograms to target gain curves are the following:

NAL-RP FITTING FORMULA CALCULATION

The insertion gains at each frequency are calculated according to
the following formulas:

Freq (Hz)

Insertion Gain (dB)

250

X + .31 HTL - 17dB + S

500

X + .31 HTL - 8dB + S

750

X + .31 HTL - 3dB + S

1000

X + .31 HTL + 1dB + S

1500

X + .31 HTL + 1dB + S

2000

X + .31 HTL - 1dB + S

3000

X + .31 HTL - 2dB + S

4000

X + .31 HTL - 2dB + S

6000

X + .31 HTL - 2dB + S

8000

X + .31 HTL - 2dB + S

Where you calculate X and S (Slope correction values) as follows:

a. T = (HTL @ 500Hz) + (HTL @ 1000Hz) + (HTL @ 2000Hz)
b. If T is less than or equal to 180dB, then

X = .05 x T

If T is greater than 180dB, then

X = (.05 x 180dB) + (.116 x ( T - 180dB)

c. If the HTL at 2000 Hz is less than 95dB, S = 0
If the HTL at 2000 Hz is greater than or equal to 95dB, then

apply the following slope correction factors at each frequency
based on the 2000Hz measurement:

2000Hz

250 500 750 1000 1500 2k 3k 4k 6k 8k

HTL (dB)

95

4

3

1

0

-1

-2

-2 -2

-2 -2

100

6

4

2

0

-2 -3

-3 -3

-3 -3

105

8

5

2

0

-3

-5

-5 -5

-5 -5

110

11

7

3

0

-3

-6

-6 -6

-6 -6

115

13

8

4

0

-4

-8

-8 -8

-8 -8

120

15

9

4

0

-5

-9

-9 -9

-9 -9