Frye FONIX 7000 User Manual

Page 38

Advertising
background image

30

FONIX 7000 Hearing Aid Analyzer

2.2.3 Setting the user mode

There are two different user modes available on the 7000 test system: Beginner
and Expert. In the Beginner mode, extra help windows pop up automatically
when function keys are pressed in order to give the user more direction in using
the instrument. In Expert mode, those help windows are eliminated.

To change the user mode:
1. Press [MENU] in the Opening screen to open the Setup menu.
2. Use [∨, ∧] to select User under Other.
3. Use [<, >] to choose EXPERT or BEGINNER.
4. Press [EXIT] to return to the Opening screen. The new user mode will be in

effect.

2.2.4 Explaining the settings

The following is an explanation of all the various settings in the Setup menu.

General
PRINT LABEL: Status of the printing label. Choose ON or OFF.

PRINTER: Printer used for printouts. Choose INTERNAL or EXTERNAL. See
Section 2.4.

EXT PRNT COLOR: Color used when an external printer is selected. Choose
BLACK & WHITE or COLOR.

EXT PRNT LANGUAGE: Language used when the external printer is selected.
Choose HPCL or EPSON.

EXT PRNT #SCRNS/PAGE: The number of screens printed on a page using an
external printer. See Section 2.4.7.

RS232 BAUDRATE: Baud rate of the RS232 connection. This is the speed of the
connection of the analyzer with a personal computer. The Windows-based pro-
gram used on the computer end will need to have the same baud rate selection.

Other
AID TYPE: Type of hearing aid being measured. Choose between STANDARD
(linear), AGC, and ADAPTIVE AGC. This will affect the delays of pure-tone
sweeps as well as some other measurements. See Section 2.3.1.3 for details.

GRAPH SCALING: Scaling used when making measurements. NORMAL scales
to the upper-most curve displayed. AUTO scales to the current selected curve.

USER: The selected user mode. See Section 2.2.3 for details.