Pogo, Berger – Frye FONIX 7000 User Manual

Page 198

Advertising
background image

190

FONIX 7000 Hearing Aid Analyzer

POGO

Prescription of Gain/Output (POGO) for Hearing Aids
by: Geary A. McCandless, PhD,

& Poul Erik Lyregaard, MSc

Hearing Instruments vol 34 #1, 1983

Required insertion gain is calculated as follows:

Frequency (Hz) Insertion Gain (dB)

250

1/2 HTL - 10 dB

500

1/2 HTL - 5 dB

*

750

1/2 HTL - 2.5 dB

1000

1/2 HTL

* 1500

1/2 HTL

2000

1/2 HTL

3000

1/2 HTL

4000

1/2 HTL

* 6000

1/2 HTL

* 8000

1/2 HTL

Note: Frequencies preceded by an asterisk (*) are
interpolated because the article does not reference them.

BERGER

A Method of Hearing Aid Prescription
by: Kenneth W. Berger, PhD, Eric N. Hagberg, M.A.,
& Robert L. Rane, PhD
Hearing Instruments July 1978

Required insertion gain is calculated as follows:

Frequency (Hz)

Insertion Gain (dB)

*

250

1/2 HTL

500

1/2 HTL

*

750

1/2 HTL

1000

1/1.6 HTL

*

1500

1/1.5 HTL

2000

1/1.5 HTL

3000

1/1.7 HTL

4000

1/2 HTL

*

6000

1/2 HTL

*

8000

1/2 HTL

Note: Frequencies preceded by an asterisk (*) are
interpolated because the article does not reference them.

Advertising