Real-Ear Measurements

155

• From 3000 Hz–8000 Hz, measured data points from seven consecutive

frequencies 100 Hz apart are averaged to get the curve point at the center
frequency.

In normal pure-tone log smoothing:

• Each center frequency is smoothed by averaging it together with the out-

puts from the next upper and lower 12th octave frequency.

To turn on/off log smoothing:

1. Enter the measurement screen.
1. Press [MENU] to enter the local menu.
2. Use [∨, ∧] to select Smoothing under Display.
3. Use [<, >] to choose LOG or OFF.
4. Press [EXIT] to return to the measurement screen. All measurements will

have the chosen smoothing status.

Figure 6.4.11—A comparison of a smoothed (top) and unsmoothed (bottom) curve.