Automated Test Sequences

119

Figure 5.2.3—ANSI 92 test results

5.2.4 Viewing ANSI measurements in the Coupler screen

It’s possible to copy ANSI response curves to the Coupler screen where they
can be compared to other response curves. To do this:

1. Run the ANSI 92 test sequence.
2. Press [F6].
3. Use [∨] to select Store Crvs Coup, if necessary.
4. Press [>] to store the curves and close the menu. The following is a list of

where the curves are stored in the Coupler screen. Any previous measure-
ments stored in those curve locations will be overwritten.
• 50 dB SPL—CRV3
• 60 dB SPL—CRV4
• 70 dB SPL—CRV5
• 80 dB SPL—CRV6
• 90 dB SPL—CRV7

5. Press [EXIT] and [F1] to enter the Coupler screen and view the ANSI

response curves.