6 directional hearing aids – Frye FONIX 7000 User Manual

Page 168

Advertising
background image

160

FONIX 7000 Hearing Aid Analyzer

5. Press [START] to measure the frequency response of the hearing aid. Press

[STOP] when the measurement has stabilized.

6. Use [F2] to select the REAR 3.
7. Use [F5] to select Composite.
8. Use [∨, ∧] to set the source amplitude to 65 dB SPL.
9. Press [START] to start the measurement. After the measurement has sta-

bilized, press [STOP] to stop the measurement. Figure 6.5.4 illustrates an
example of a digital hearing aid with noise suppresion. Curve 2 shows the
response of the aid to the Digital Speech curve, and Curve 3 shows the
response of the aid to the Composite signal. .

Figure 6.5.4—Testing noise suppression of a digital hearing aid. REAR 2 was tested using
Digital Speech. REAR 3 was tested using the Composite Signal.

6.6 Directional Hearing Aids

The directional advantage provided by a hearing aid can be easily measured
with real-ear measurements. This section describes a method of measuring
directionality using the Real-ear Insertion Gain screen.

6.6.1 Using the insertion gain technique

The Real-ear Insertion Gain screen is ideal for performing real-ear measure-
ments of hearing aid directionality.

To test directionality:
1. Set up the patient for real-ear measurements by inserting the probe micro-

phone and hearing aid.

2. Place the sound field speaker behind the patient at 180° or 135° azimuth.

Point the reference microphone of the probe microphone set towards the
sound field speaker.

3. Enter the Real-ear Insertion Gain screen.