3 example – Frye FONIX 7000 User Manual

Page 78

Advertising
background image

70

FONIX 7000 Hearing Aid Analyzer

Figure 3.4.2—Pure-tone measurement

3.4.3 Example

Following the instructions above, run 4 Gain curves using amplitudes 50, 60,
70, and 80 dB SPL. Use a signal source of COMPOSITE.

This scenario will provide you with a family of curves. Whenever the curves sit
on top of each other, the gain of the aid is the same for the amplitudes of those
curves. This indicates the aid is running linearly. If the curves provide progres-
sively less gain as the input rises, then the aid has some sort of compression
circuit. Figure 3.4.3 shows two examples of this scenario.

Here are the steps:

1. Look at the graph. If it has a display of SPL, press [MENU] and change the

DISPLAY from SPL to GAIN, using the arrow keys. Press [EXIT] to return to
the Coupler screen.

2. Use [F2] to select CRV 1. (Press [F2]. Use [∨, ∧] to select CRV1. Press [>] to

complete the selection and close the pop-up window.)

3. Use [∨, ∧] to select a source level of 50 dB SPL.
4. Use [F5] to select a source type of COMPOSITE.
5. Press [START] to run the first measurement. Press [STOP] when the mea-

surement has stabilized.

Advertising