FONIX 7000 Hearing Aid Analyzer

5. Press [F1] from the Real-Ear Navigation screen to enter the Audiogram Entry

screen.

6. Press [MENU] from the Audiogram Entry screen to enter the local menu.
7. Use [∨, ∧] to select RECD Mode under Misc, and use [<, >] to choose

Custom Measured.

8. Use [∨, ∧] to select RECD Transducer and use [<, >] to choose ER3.

Probe mic

Probe mic

ER3

INSER

T

EARPHONE

Foam plug or

custom earmold

Figure 6.3.5C—Insert probe microphone

Figure 6.3.5D—Insert foam eartip connected

to insert microphone.

9. Press [EXIT] to close the menu.
10. Use [F2] in the Audiogram Entry screen to select RECD. You may have to

push the button several times.

11. Press [START] to take the measurement. The results will appear in both

graphical and numeric format. See Figure 6.3.5.E

If you have previously taken an RECD measurement, you can input it (with-
out measuring it) by selecting RECD with [F2] and using the arrow keys as
you would to input HTL values.