SRS Labs SR850 User Manual

Page 10

Advertising
background image

SR850 DSP Lock-In Amplifier

1-8

DTRC (?) i {, j}

6-11

Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Trace to trace j (1,2,3,4).

DSCL (?) {x}

6-11

Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Range to x.

DOFF (?) {x}

6-11

Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Center value to x.

DHZS (?) {i}

6-12

Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Horizontal Scale to 2 ms (0) through
200 ks (32) per div.

RBIN? i

6-12

Query the bin number at the right edge of the Full (i=0), Top (i=1) or Bottom (i=2) chart display.

CURSOR

page description

CSEK (?) {i}

6-13

Set (Query) the active display Cursor Seek mode to Max (0), Min (1) or Mean (2).

CWID (?) {i}

6-13

Set (Query) the active display Cursor Width to Off (0), Narrow (1), Wide (2) or Spot (3).

CDIV (?) {i}

6-13

Set (Query) the active display Chart Divisions to 8 (0), 10 (1) or None (2).

CLNK (?) {i}

6-13

Set (Query) the Cursor Control Mode to Linked (0) or Separate (1).

CDSP (?) {i}

6-13

Set (Query) the active display Cursor Readout to Delay (0), Bin (1), Fsweep (2) or Time (3).

CMAX

6-13

Move active chart cursor to max or min. Same as pressing [CURSOR MAX/MIN] key.

CURS? i

6-13

Query the cursor horz,vert position of Full (0), Top (1) or Bottom (2) chart display.

CBIN (?) {i}

6-13

Set (Query) the center of the cursor region in the active chart display. i is the bin number.

MARK

page description

MARK

6-14

Places a mark in the data buffer. Same as pressing [MARK] key.

MDEL

6-14

Delete the nearest mark to the left of the cursor. Same as pressing <Marker Delete> softkey.

CNXT

6-14

Move active chart cursor to next mark to the right.

CPRV

6-14

Move active chart cursor to next mark to the left.

MACT?

6-14

Query the number of active marks. Also returns the active mark numbers.

MBIN? i

6-14

Query the bin number of mark #i.

MTXT (?) i {,s}

6-14

Set (Query) the label text for mark #i.

AUX INPUT/OUTPUT

page description

OAUX ? i

6-15

Query the value of Aux Input i (1,2,3,4).

AUXM(?) i{, j}

6-15

Set (Query) the Output Mode of Aux Output i (1,2,3,4). j selects Fixed (0), Log (1) or Linear (2).

AUXV (?) i {, x}

6-15

Set (Query) voltage of Aux Output i (1,2,3,4) to x Volts. -10.500

x

10.500. Fixed Output

Mode only.

SAUX (?) i {, x, y, z}

6-15

Set (Query) the Aux Output i (1,2,3,4) Sweep Limits to Start (x), Stop (y) and Offset (z) voltag-
es. 0.001

x,y

21.000 and -10.500

z

10.500.

TSTR (?) {i}

6-15

Set (Query) the Trigger Starts Scan? mode to No (0) or Yes (1).

MATH

page description

SMTH i

6-16

Smooth the data within the active chart using 5 (0), 11 (1), 17 (2), 21 (3), 25 (4) point width.

COPR (?) {i}

6-16

Set (Query) the Calculator Operation to +, -, x, /, sin, cos, tan,

x, x2, log, 10x (i=0...10).

CALC

6-17

Do the Calculation selected by COPR with the argument set by CTRC or CARG.

CAGT (?) {i}

6-17

Set (Query) the Calculation Argument Type to Trace (0) or Constant (1).

CTRC (?) {i}

6-17

Set (Query) the Trace Argument to Trace i (1,2,3,4).

CARG (?) {x}

6-17

Set (Query) the Constant Argument value to x.

FTYP (?) {i}

6-17

Set (Query) the Fit Type to Linear (0), Exponential (1) or Gaussian (2).

FITT i, j

6-17

Fit the data within the chart region between i% and j% from the left edge. 0

i,j

100.

PARS ? i

6-17

Query the fit parameters a (0), b (1), c (2) or t0 (3).

STAT i, j

6-17

Statistically analyze the data within the chart region between i% and j% from the left edge.
0

i,j

100.

SPAR ? i

6-17

Query the Statistical results mean (0), standard dev (1), total (2) or delta time (3).

STORE AND RECALL FILE

page description

FNAM (?) {s}

6-18

Set (Query) the current File Name to string s.

SDAT

6-18

Save the Active Display's Trace Data to the file specified by FNAM.

SASC

6-18

Save the Active Display's Trace Data in ASCII format to the file specified by FNAM.

SSET

6-18

Save the Settings to the file specified by FNAM.

Advertising