7-9

Performance Tests

3. Common Mode Rejection

This test measures the common mode rejection of the lock-in.

Setup

We will use the internal oscillator sine output to provide the signal.

Connect the Sine Out to both the A and B inputs of the lock-in. Use equal length cables from A and B to a
BNC TEE. Connect the cable from the Sine Out to the TEE. Do not use any termination.

Procedure

1) {PRESET} (Turn the lock-in off and on with the [

] key pressed)

2) Press the keys in the following sequence:

[REF/PHASE]

<Ref. Frequency>

[1] [0] [0] [ENTER]

[DISPLAY/SCALE]

<Type and Trace>

Highlight the trace number.
Use the knob to select trace 3 (R) for the top display.

[INPUT/FILTER]

<Coupling>

Select DC.

3) The value of R should be 1.000 V (within 2%).

4) Press

[INPUT/FILTER]

<Source>

Select A-B.

[GAIN/PHASE]

<Sensitivity>

Use the knob to select 200 µV.

5) Record the value of R.

6) This completes the CMRR measurement test. The common mode rejection is 20log(1.0/R) where R is

in Volts. Enter the results of this test in the test record at the end of this section.