11 remote programming display and scale commands – SRS Labs SR850 User Manual

Page 187

Advertising
background image

6-11

Remote Programming

DISPLAY and SCALE COMMANDS

ASCL

The ASCL command auto scales the active display. This command is
just like pressing the [AUTO SCALE] key. Only Bar and Chart displays
are affected.

ADSP (?) {i}

The ADSP command selects the active display. The parameter i selects
Full (i=0), Top (i=1) or Bottom (i=2). The selected display must be pres-
ently displayed on the screen otherwise an error will result.

SMOD (?) {i}

The SMOD command sets or queries the screen format. The parameter i
selects Single or full screen display (i=0), or Up/Down dual display (i=1).

MNTR (?) {i}

The MNTR command sets or queries the monitor display mode. The
parameter i selects Settings monitor (i=0), or Input/Output monitor (i=1).

DTYP (?) i {, j}

The DTYP command sets or queries the display type. The parameter i
selects the Full (i=0), Top (i=1) or Bottom (i=2) display and is required.
The parameter j selects Polar (i=0), Blank (i=1), Bar (i=2) or Chart (i=3).
An error is generated if the DTYP command tries to set the display type
of a display which is not on the screen. The DTYP?i query can check any
display's type.

DTRC (?) i {, j}

The DTRC command sets or queries the displayed trace number. The
parameter i selects the Full (i=0), Top (i=1) or Bottom (i=2) display and is
required. The parameter j is the trace number (1, 2, 3 or 4). An error is
generated if the DTRC command tries to set the trace number of a dis-
play which is not on the screen. The DTRC?i query can check any dis-
play's trace.

DSCL (?) i {, x}

The DSCL command sets or queries the display range. The parameter i
selects the Full (i=0), Top (i=1) or Bottom (i=2) display and is required.
The parameter x is the range (real number with the units of the displayed
trace). The value of x is limited to 10

-18

< x < 10

18

. An error is generated

if the DSCL command tries to set the range of a display which is not on
the screen. The range only affects bar and chart display types.

DOFF (?) i {, x}

The DOFF command sets or queries the display center value or offset.
The parameter i selects the Full (i=0), Top (i=1) or Bottom (i=2) display
and is required. The parameter x is the center value (real number with
the units of the display trace). The value of x is limited to 10

-12

< x <

10

12

. An error is generated if the DOFF command tries to set the center

of a display which is not on the screen. The center only affects bar and
chart display types.