7 sr850 service – SRS Labs SR850 User Manual

Page 253

Advertising
background image

8-7

SR850 Service

Press

[AUTO PHASE]

The bottom display should read 0.0°.

7. Press

[INPUT/FILTERS]

<Line Notches> to select Line

[GAIN/TC]

<Sensitivity>

Rotate the knob to select 50 mV full scale

Adjust the Line Notch Depth pot until R (top display) is between 20.0 and 25.0 mV.

Then adjust the Line Notch Freq pot to minimize the reading of R AND until

θ

(bottom display) is

EITHER 0° or 180° (within 1°).

8. Now adjust the Line Notch Depth pot until R is less than 1.0 mV

9. Press

[INPUT/FILTERS]

<Line Notches> three times to select Out

[GAIN/TC]

<Sensitivity>

Rotate the knob to select 1V full scale

[REF/PHASE]

<Ref. Frequency> Select Reference Frequency

[1] [2] [0] [ENTER] Enter 120 Hz OR
[1] [0] [0] [ENTER] Enter 100 Hz

After the readings are stable, the top display should read 1.00 V (±2%).

Press

[AUTO PHASE]

The bottom display should read 0.0°.

10. Press

[INPUT/FILTERS]

<Line Notches> twice to select 2xLine

[GAIN/TC]

<Sensitivity>

Rotate the knob to select 50 mV full scale

Adjust the 2xLine Notch Depth pot until R (top display) is between 20.0 and 25.0 mV.

Then adjust the 2xLine Notch Freq pot to minimize the reading of R AND until

θ

(bottom display)

is EITHER 0° or 180° (within 1°).

11. Now adjust the 2xLine Notch Depth pot until R is less than 1.0 mV

12. This completes the notch filter adjustment procedure.

13. Turn the unit OFF. Replace the top lid. The lid slides forward and down to engage the clip at the

center front. Replace the four black screws to secure the lid.