background image

4-3

Front Panel

Ch1 & Ch2 Outputs
The Channel 1 and Channel 2 outputs can be con-
figured to output a voltage from -10 V to +10 V
proportional to X, Y, R,

θ

, or Traces 1 through 4.

±10 V is full scale. The outputs can source 10 mA
maximum.

Signal Inputs
The input mode may be single-ended, A, or diffe-
rential, A-B. The A and B inputs are voltage inputs
with 10 M

, 25 pF input impedance. Their connec-

tor shields are isolated from the chassis by 10

(Ground) or 1 k

(Float). Do not apply more than

50 V to either input. The shields should never
exceed 1 V. The I (current) input is 1 k

to a virtu-

al ground.