Remote Programming

INPUT and FILTER COMMANDS

ISRC (?) {i}

The ISRC command sets or queries the input configuration. The parame-
ter i selects A (i=0), A-B (i=1) or I (i=2).

IGAN (?) {i}

The IGAN command sets or queries the conversion gain of the current
input. The parameter i selects 1 M

(i=0), 100 M

(i=1). This parameter

has no effect unless the input is configured to measure current.

Changing the current gain does not change the instrument sensitivity.
Sensitivities above 10 nA require a current gain of 1 M

. Sensitivities

between 20 nA and 1 µA automatically select the 1 M

current gain. At

sensitivities below 20 nA, changing the sensitivity does not change the
current gain.

IGND (?) {i}

The IGND command sets or queries the input shield grounding. The
parameter i selects Float (i=0) or Ground (i=1).

ICPL (?) {i}

The ICPL command sets or queries the input coupling. The parameter i
selects AC (i=0) or DC (i=1).

ILIN (?) {i}

The ILIN command sets or queries the input line notch filter status. The
parameter i selects Out or no filters (i=0), Line notch in (i=1), 2xLine
notch in (i=2) or Both notch filters in (i=3).