Sine output amplitude accuracy and flatness, 19 performance tests, Setup – SRS Labs SR850 User Manual

Page 237: Procedure

Advertising
background image

7-19

Performance Tests

8. Sine Output Amplitude Accuracy and Flatness

This test measures the amplitude accuracy and frequency response of the internal oscillator Sine Out.

Setup

We will use the lock-in to measure the Sine Out. Connect the Sine Out to the A input of the lock-in.

Procedure

1) {PRESET} (Turn the lock-in off and on with the [

] key pressed)

2) Press the keys in the following sequence:

[DISPLAY/SCALE]

<Type and Trace>

Highlight the trace number.
Use the knob to select trace 3 (R) for the top display.

3) Amplitude accuracy is verified at 1 kHz using various sensitivities. For each sine amplitude and sensi-

tivity setting in the table below, perform steps 3a through 3b.

Sensitivity

Sine Output Amplitude

1 V

1.000 Vrms

200 mV

0.200 Vrms

50 mV

0.050 Vrms

10 mV

0.010 Vrms

a) Press

[REF/PHASE]

<Sine Output>

Enter the sine amplitude from the table.

[GAIN/TC]

<Sensitivity>

Use the knob to select the sensitivity from the table.

b) Wait for the R reading to stabilize. Record the value of R.

c) Repeat 3a and 3b for each amplitude in the table.

4) Frequency response is checked at frequencies above 1 kHz. The sine amplitude is set to 1 Vrms for

all frequencies. The test frequencies are listed below.

Test Frequencies

24 kHz
48 kHz
72 kHz
96 kHz